product description
The Gentier 48E / 48R Automatic PCR Analysis System is Tianlong Technology specializes in tailor-made by SMB, mobile laboratory, on-site testing, supports absolute quantitative, relative quantitative, isothermal amplification, melting curve, genotyping , End-point fluorescence analysis and other experimental functions. With excellent performance and portability, your test is more convenient, accurate, and efficient!
Instrument characteristics
● The lift is rapid, the temperature rise rate is ≥ 8.0 ° C / s, the cooling rate is ≥ 6.2 ° C / s;
● Highlight the LED light source, with 4/2 fluorescence detection channels, complete fluorescence scanning of all hole positions only need to be 2s;
● 7-inch full color touch screen, the device can be run independently, store more than 1000 experimental data;
● No computer, you can directly analyze the experimental data on the instrument;
● Automatically recover the experimental function when it is turned off, and there is no need to wait for PC and software to open;
● Small size, light weight, easy to carry, no calibration work is required after moving.
Product parameters
Equipment model | Gentier 48E | Gentier 48R |
Fluorescence channel | 4 | 2 |
Sample flux | 48 | |
Suitable fluorescein | Channel 1: Fam, Sybr Green I, SYTO 9, Evagreen, LC Green Channel 2: HEX, VIC, TET, JOE Channel 3: ROX, Texas RED Channel 4: CY5 | Channel 1: Fam, Sybr Green I, SYTO 9, Evagreen, LC Green Channel 2: HEX, VIC, TET, JOE |
light source | High Light Long Life Maintenance LED Light Source | |
Detector | Photoelectric diode (PD) | |
Detection method | All fluorescent channels simultaneously scan, no fluorescence edge effect | |
Detection time | Complete all of the 48 hole positions of all fluorescence channels in 2 seconds | |
Lifting temperature rate | Earthquan rate ≥ 8.0 ° C / s Cooling rate ≥ 6.2 ° C / s | |
Temperature accuracy | ≤0.1 ° C | |
Temperature control program | Support standard PCR, gradient PCR, landing PCR, long fragment PCR, etc. | |
Analysis function | Qualitative analysis, absolute quantitative analysis, relative quantitative analysis, end point fluorescence analysis, melting curve analysis, SNP analysis, etc. | |
Sample linearity and repeatability | Linear: /R/≥0.999 Repeatability: CT value CV≤0.5% | |
Manifest | Single machine operation: 7-inch touch screen control operation; Network Operation: Control through PC-side software in direct or LAN. | |
Applicable consumables | 0.2ml transparent skirt 48-well plate, 8 interlock, single tube | |
Power off protection | The experiment automatically restores the function of the operation when power is powered off, and there is no need to wait for PC and software to open. | |
data storage | Single machine runs more than 1000 experimental data files | |
Report custom function | Pre-stored various industry experiment report templates All open universal report features, users can customize report content and form | |
LIS function | Can export formats such as CSV, Excel, TXT Open data port, synchronous support and LIS system interconnection | |
Single machine analysis results | No computer, you can directly analyze the number of experiments on the instrument, connect special thermal printers (optional) can directly output sample amplification curves and CT values |
Tianlong Technology is based on market-oriented, and has been based on the research and development of science and technology, and pursues independent brands.